考慮到臨床工作對(duì)椎間孔鏡的性能有較高要求,再加上脊柱內(nèi)窺鏡本身性質(zhì)特殊,所以掌握椎間孔鏡的日常清潔保養(yǎng)、功能檢查及處理措施等知識(shí),就是非常重要的。
Considering that clinical work has high requirements for the performance of intervertebral foramen scope, and the special nature of spinal endoscope itself, it is very important to master the knowledge of daily cleaning and maintenance, functional examination and treatment measures of intervertebral foramen scope.
第一步:使用軟毛巾或紗布在流動(dòng)自來水下擦拭表面,去除污漬和血跡。使用大號(hào)和小號(hào)軟毛刷分別清潔工作通道和左右灌洗通道。
Step 1: wipe the surface with a soft towel or gauze under running tap water to remove stains and blood stains. Use large and small soft bristles to clean the working channel and the left and right irrigation channels respectively.
第二步:使用大號(hào)注射器或高壓水槍清潔灌洗通道和工作通道。
Step 2: use a large syringe or high-pressure water gun to clean the irrigation channel and working channel.
第三步:將椎間孔鏡置于浸泡箱,在去污漬生物酶稀釋液內(nèi)浸泡5min;然后放入3%的過氧化氫消毒劑中浸泡5min。
Step 3: put the intervertebral foramen scope into the soaking box and soak it in the stain removing biological
enzyme diluent for 5min; Then soak it in 3% hydrogen peroxide disinfectant for 5min.
第四步:重復(fù)第一步和第二步。確保內(nèi)外表面和腔道清潔光滑,無血污;物鏡和目鏡徹底清潔。
Step 4: repeat step 1 and step 2. Ensure that the internal and external surfaces and cavities are clean and smooth without blood stains; The objective lens and eyepiece shall be thoroughly cleaned.
第五步:用去離子水(蒸餾水、反滲水等)再次沖洗椎間孔鏡的內(nèi)外表面和腔道(如缺少此步驟,滅菌后將在此內(nèi)外表面殘留水漬)
Step 5: use deionized water (distilled water, reverse osmosis water, etc.) to wash the inner and outer surfaces and lumens of the intervertebral foramen mirror again (if this step is missing, water stains will remain on the inner and outer surfaces after sterilization)
第六步:將椎間孔鏡置于干燥箱烘干(不超過60°)(如缺少此步驟,等離子滅菌時(shí)可能造成破壞;高溫滅菌后可能殘留水漬)
Step 6: put the intervertebral foramen scope in the drying oven for drying (no more than 60 °) (if this step is missing, damage may be caused during plasma sterilization; water stains may remain after high-temperature sterilization)
第七步:將醫(yī)用潤滑劑涂抹于干紗布,擦拭椎間孔鏡的外表面(物鏡和目鏡除外);用注射器噴涂少許醫(yī)用潤滑劑于椎間孔鏡的內(nèi)部腔道。
Step 7: apply the medical lubricant to the dry gauze and wipe the outer surface of the intervertebral foramen lens (except the objective lens and eyepiece); Use a syringe to spray a little medical lubricant on the internal cavity of the intervertebral foramen mirror.
內(nèi)窺鏡設(shè)備維修人員推薦您要重視清潔保養(yǎng)的情況,會(huì)直接影響到椎間孔鏡的耐用性,大家要認(rèn)真對(duì)待,認(rèn)真落實(shí)!
The endoscope equipment maintenance personnel recommend that you pay attention to the cleaning and maintenance, which will directly affect the durability of the intervertebral foramen scope. Everyone should take it seriously and implement it carefully!
以上就是今天為大家介紹的關(guān)于問題以及注意事項(xiàng)的全部內(nèi)容,希望對(duì)大家能有所幫助,想要了解更多內(nèi)容可以電話聯(lián)系我們或者關(guān)注我們的網(wǎng)站
http://483177.com,我們的工作人員會(huì)及時(shí)與您取得聯(lián)系。
The above is all about the problems and precautions introduced to you today. I hope it can help you. If you want to know more, you can contact us by phone or follow our website http://483177.com , our staff will contact you in time.